8-3 Operating procedures

} Be sure to turn ON the power of the control box for microscope or that of the microscope frame before turning ON
the power of D2-CB. If you turn ON the power in the wrong order, D2-CB does not recognize the control box for
microscope or the microscope frame correctly.
Automatic scale change
When the objective magnification is changed by changing the revolving nosepiece of the microscope, the scale size
is changed automatically according to the magnification changed.
To use this function, the objective magnification must be set to DP2-SAL in advance. Set the objective magnification
according to the microscope setting wizard displayed when the power is turned ON for the first time. If DP2-SAL has
already been started, refer to "Setting the microscope (P.78)" and set the objective magnification.
a
c
Hand switch for exposure U-HSEXP
U-HSEXP can be attached to the microscope by connecting U-CBS to U-HSEXP with the cable. With this
operation, you can acquire images while operating the microscope.
U-HSEXP is a same function as the [Snap] button on the control window. Pressing the U-HSEXP switch allows you
to acquire and save images.
} For procedures to connect to U-CBS, refer to P. 90.
} For procedure to attach to the microscope, refer to P. 91.
} You cannot record movies with U-HSEXP. Use the [Movie] button in the control window . (P.36)
92
Displaying scale
1
b
2
} For procedures to display scales, refer to "Setting scale display (P.45)".
Select the [Menu] tab a and select the [Display] tab b in the tab
displayed.
Press the [ON] button of [Scale] c to display the scale.