Advanced Settings; Setting Ae Area; Shading Correction - Olympus DP2-SAL Instructions Manual

Standalone connection kit
Table of Contents

Advertisement

4-4 Advanced settings

b
b
d
e
: The corrected image is not acquired yet.
:
The corrected image has been acquired.
40
a
c
a
c
1

Setting AE area

You can specify whether or not to display the metering area on the
display. The metering area is shown with the yellow line on the display.
For setting the metering area, refer to P.33.
1
Select the [Menu] tab a and select the [Camera Control] tab b in
the tab displayed.
2
Press [ON] button or [OFF] button of [Display AE area] c to select
Show or Hide.
2

Shading correction

The unevenness of brightness caused by the uneven illumination can
be corrected to the even brightness.
To use the shading correction function, the objective
CAUTION
magnification and the zoom/mag.changer magnification
of the microscope must be specified in advance. For
details, refer to "Setting the microscope (P.78)".
Acquisition of corrected image
The shading correction image is acquired per each combination
of the specified objective magnification and zoom/mag.changer
magnification. It takes approximately 1 minute to acquire one
combination.
Before acquiring the corrected image, place the specimen to bring it
into focus and remove the specimen in advance.
· When acquiring the shading correction image, if there is
CAUTION
dust or dirt in the acquisition area, the image may not be
acquired properly. Be sure to confirm that there is no dust
or dirt in the acquisition area.
· Up to 28 combinations of corrected images can be acquired.
If you attempt to acquire more than 28 combinations of
corrected images, an error message appears. In this case,
set the microscope again. For details, refer to "Setting the
microscope (P.78)".
1
Select the [Menu] tab a and select the [Quality] tab b in the tab
displayed.
2
Select the [Calibration] button c . The [Calibration] wizard is displayed.
3
Use [Select objective for calibration] d and [Select Manual Zoom/
Magnification Changer for calibration] e to select the objectives
and the zoom/magnification changer you want to correct, and
select [Next].