8 Hygienic reprocessing
Result

8.10 Preparing parts for reuse

Requirement
Result
104
EN
MEDUVENT Standard
2. Steam-sterilize parts using a device conforming to EN 285.
When doing so:
Use a temperature of 134 °C and a dwell time of 5 minutes
or
Use a temperature of 132 °C and a dwell time of 4 minutes
The instructions of the manufacturer of the steam sterilizer
must be followed.
The parts have been steam-sterilized.
The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage resulting from use
(e.g. tension cracks or cable breaks).
2. Replace damaged parts.
3. Assemble the reusable breathing circuit
the reusable breathing circuit", page
4. Install the accessories.
5. Reconnect the power supply
supply", page
34).
6. Perform a function check
7. Place parts in storage in accordance with the storage
conditions
(see "15 Technical data", page
The parts are ready for use again.
(see "8.3.2 Assembling
90).
(see "4.2 Connecting a power
(see "5 Function check", page
138).
44).