Preparing Parts For Reuse - Weinmann LIFE-BASE light XS Instructions For Use Manual

Portable unit
Table of Contents

Advertisement

8 Hygienic reprocessing
Result

8.6 Preparing parts for reuse

Requirement
Result
154
EN
LIFE-BASE
3. Brush the soiled areas for at least 2 minutes under running
faucet water using a cleaning brush until they look clean.
4. Rinse the cleaned areas for 2 minutes under running
faucet water.
5. If necessary: Open the bags.
6. Wash parts with cleaning agent and disinfectant in
the washing machine without spinning:
Pre-washing: 5 min. at 20°C
Main washing: 30 min. at 30°C
The parts are mechanically reprocessed.
The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage due to use, e.g., cable breakage.
2. Replace damaged parts.
3. If available: Attach the belts to the portable unit
(see "4.3 Attaching the belt to the portable unit", page
4. Attach the devices to the portable unit
devices to the portable unit", page
5. If available: Attach the protective bags and accessories bags
to the portable unit
6. If available: Attach the oxygen cylinder to the portable unit
(see "4.4 Attaching the oxygen cylinder to the portable unit",
page
38).
7. Connect the accessories to the device.
8. Perform a function check
9. With longer storage period: Place parts in storage in
accordance with the storage conditions
page
158).
The parts are ready for use again.
(see "4.5 Attaching the
43).
(see "4 Assembly", page
(see "5 Function check", page
36).
25).
81).
(see "10 Storage",

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Life-base 1 ngLife-base iiiLife-base lightLife-base 1 ng xlLife-base 3 ngLife-base 1 ng xs ... Show all

Table of Contents