Technical Data; Input; Performance Characteristics - Endress+Hauser Memosens Wave CKI50 Operating Instructions Manual

Process spectrometer for the measurement of color
Table of Contents

Advertisement

Technical data

Measured variable
Measuring range
Reference conditions
Repeatability
Long-term reliability
1)
L*a*b* color model defined by the International Commission of Illumination
38
12
Technical data
12.1

Input

1)
CIE L*a*b*
, observer 2°, illuminant D65, as per DIN EN ISO 11664-4
CIE L*a*b* is a device-independent color space consisting of 3 color values:
• L* lightness axis
Specified operational range: 0 to 100
• a* green-red axis
Specified operational range: –150 to 100
• b* yellow-blue axis
Specified operational range: –100 to 150
Application
Halogen wavelength range
CIE L*a*b*
12.2

Performance characteristics

25 °C (77 °F), 1013 hPa (15 psi), heat-up time 5 hours
Repeatability
L*
a*
b*
1)
Measured according to DIN ISO 15839 with liquid color solutions in the range L* from 60 to 100, a* from
-47 to 85; b* from -44 to 98
Drift per 24h
L*
a*
b*
1)
Measured according to DIN ISO 15839 with liquid color solutions in the range L* from 60 to 100, a* from
-47 to 85; b* from -44 to 98
Regular referencing ensures the drift is largely compensated.
Specified operational range
380 to 830 nm
L* = 0 to 100
a* = -150 to 100
b* = -100 to 150
Specified operational range
0 to 100
-150 to 100
-100 to 150
Specified operational range
0 to 100
-150 to 100
-100 to 150
Memosens Wave CKI50
Repeatability
< 0.1 % of the span of the specified
1)
operational range
Drift per 24h
< 0.03 % of the span of the
1)
specified operational range
Endress+Hauser