Technical data
Electrical connection
Potential equalization
Terminals
Cable entries
Cable specification
Reference operating
conditions
Maximum measurement
error
104
→  51
→  54
Spring terminals for wire cross-sections 0.5 to 2.5 mm
Cable entry thread
• NPT ½"
• G ½"
Cable gland
• For standard cable: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• For armored cable: M20 × 1.5 with cable 9.5 to 16 mm (0.37 to 0.63 in)
If metal cable entries are used, use a grounding plate.
→  44
16.6

Performance characteristics

• Error limits following DIN EN 29104, in future ISO 20456
• Water, typically: +15 to +45 °C (+59 to +113 °F); 0.5 to 7 bar (73 to 101 psi)
• Data as indicated in the calibration protocol
• Accuracy based on accredited calibration rigs according to ISO 17025
Error limits under reference operating conditions
o.r. = of reading
Volume flow
±0.5 % o.r. ± 2 mm/s (0.08 in/s)
Fluctuations in the supply voltage do not have any effect within the specified range.
[%]
2.5
2.0
1.5
1.0
0.5
0
0
1
2
0
5
 16
Maximum measurement error in % o.r.
Electrical conductivity
Maximum measurement error not specified.
2
4
6
10
15
20
Proline Promag W 800
(20 to 14 AWG)
8
10 [m/s]
25
30
32 [ft/s]
Endress+Hauser
v
A0039745