Removal Of The Rack From The Rotator; Removal Of The Rack From The Gripper Of The Transport Arm Above The Rotator; Removal Of A Rack From The Transfer Station Of The Histocore Spectra St - Leica HistoCore SPECTRA CV Instructions For Use Manual

Coverslipper
Table of Contents

Advertisement

8
Malfunctions and Troubleshooting

8.3.5 Removal of the rack from the rotator

1. Moving the unload drawer into removal position (rescue position).
2. Insert your hand carefully into the instrument and remove the rack
rotator (→ Fig. 123-2).

8.3.6 Removal of the rack from the gripper of the transport arm above the rotator

1. Moving the unload drawer into removal position (rescue position).
2. Insert your hand carefully into the instrument and remove the rack from the gripper of the
transport arm. To do so, grasp the handle and pull it forward, along with the rack, toward the
load drawer.

8.3.7 Removal of a rack from the transfer station of the HistoCore SPECTRA ST

L If a malfunction occurs in workstation mode while the transfer device of the
HistoCore SPECTRA ST transfers a rack to the HistoCore SPECTRA CV using the transfer station,
the user must check where the rack is located.
1. Open the hood of the HistoCore SPECTRA ST.
2. Check whether the rack is still visible from the transfer station (→ Fig. 124-1).
150
Fig. 123
1
2
(→ Fig. 123-1)
from the
Version 2.1, Revision K