Cleaning and Maintenance
Note
The needle has a notch (→ Fig. 87-3) that fits exactly in the holder. The Attention symbol
(→ Fig. 87-4)
on the holder
inserting the needle into the holder. The needle must be inserted straight and all the way in order to
make sure that no negative impact on the samples arises during processing.
4. Repeat steps 1-3 for the second coverslip line (L2).
5. Finally, close the hood.
 After checking the fill level, the instrument is ready to operate and can begin processing.

7.3.2 Extended Prime

Note
• The
Extended Prime
medium bottle is replaced
respective coverslip line
• During an Extended Prime process, mounting medium is pumped through the entire hose system to
ensure that no air remains in the hose system as a result of the replacement.
• After the replacement of the mounting medium bottle, the
(→ Fig. 98-4)
1. To change the mounting medium bottle, proceed the same way as in
Changing the mounting medium bottle)
2. Take note of the information message that follows
button.
3. After the steps under
Extended Prime
122
(→ Fig. 87-2)
(→ Fig. 98-3)
rinsing program must be carried out after each time a mounting
(→ P. 78 – 6.3.1 Changing the mounting medium bottle)
L1
or L2.
buttons are inactive (grayed out).
Fig. 99
(→ P. 118 – Priming preparation)
(→ Fig. 98-3)
button.
indicates to the user that utmost care is required when
Quick Prime
and close the hood.
(→ Fig. 99)
have been completed, press the
for the
(→ Fig. 98-2)
and
(→ P. 78 – 6.3.1
and acknowledge with the
Version 2.1, Revision K
Clean
Ok