7.2.4 Interior cleaning

Warning
• There is a danger of cutting injuries during this cleaning step. Therefore, proceed with the
necessary caution and wear cut-resistant gloves
• The warning messages in
handling solvents!
1. Open the load and unload drawers and check the interior space behind the drawers for broken
glass and residual mountant.
Note
Remove all existing racks in the load and unload drawers before cleaning (→ Fig. 83).
2. Carefully remove any dirt remnants.
3. Pull the retaining lever for the unload drawer
open the drawer as far as it will move.
HistoCore SPECTRA CV
(→ P. 32 – 4.6 Refilling consumables)
Fig. 83
Cleaning and Maintenance
(→ P. 18 – 3.1 Standard delivery – packing
must be observed when
(→ Fig. 84-1)
upwards while slowly continuing to
7
list)!
107