Note
The individual rinsing programs are clarified in the following, using the coverslip line L1
The same procedures also apply for the coverslip line L2.

7.3.1 Quick Prime

Note
At the start of operation, the user is prompted after starting the instrument to carry out the
Prime
(→ Fig. 98-2) rinsing program. This step is necessary to ensure the flow rate of the mounting
medium through the needle system. The respective coverslip line is ready to operate only after Quick
Prime.
1. After the steps under
Prime
(→ Fig. 98-2)
2. Priming lasts approx. 35 seconds and uses approx. 2 ml of mounting medium. This quantity is
taken into account by the CMS.
3. After the rinsing process has been completed, reinsert the needle unit into the holder
(→ Fig. 87) and confirm the corresponding information message with OK.
HistoCore SPECTRA CV
2
3
4
Fig. 98
(→ P. 118 – Priming preparation)
button.
Cleaning and Maintenance
1
have been completed, press the
7
(→ Fig. 98-1).
Quick
Quick
121