Cleaning and Maintenance

7.2.2 TFT touchscreen

» Clean the screen using a lint-free cloth. A suitable screen cleaner can be used in compliance with
the manufacturer information.

7.2.3 Load and unload drawers

Warning
• To avoid an overflowing of a filled reagents, do not open the load drawer suddenly.
• Wear suitable protective clothing (lab coat, safety goggles, cut-resistant gloves).
1. Remove reagent vessels
instrument.
2. Check the xylene in the reagent vessels, correct the level or replace the xylene if necessary
(→ P. 32 – 4.6 Refilling consumables).
3.
Check the inside of the load drawer for reagent residue and clean up any you find (→ P. 107 –
7.2.4 Interior
4. Finally, insert the reagent vessels into the correct positions again.
L Observe the existing labeling
Fig. 82
Note
Reagent residue can also be present in the unload drawer from transporting the racks from the
coverslip line to the oven. Therefore, check the inside of the unload drawer for reagent residue and
clean if necessary
106
(→ Fig. 82-1)
cleaning).
(→ Fig. 82-2)
(→ P. 107 – 7.2.4 Interior
from the load drawer and store them outside the
for stations in the drawers.
1
cleaning).
2
Version 2.1, Revision K