7.2.9 Removing the complete unit of the needle cleaning container

The needle cleaning container is inserted into a red complete unit (→ Fig. 88-1), which can be removed
by the user through the opening in the unload drawer. The complete unit is hooked into the coverslip
line by two lateral pins (→ Fig. 88-3). The removal may be necessary if the needle cleaning container
(→ Fig. 88-2)
is stuck due to dried mounting medium and cannot be removed as described
6.3.2 Monitoring and refilling of the needle cleaning container).
Warning
• There is a danger of cutting injuries when removing the complete unit. Therefore, proceed with
the necessary level of caution and wear cut-resistant gloves
packing
list).
• The warning messages in
handling solvents!
1. Open the hood.
2. Call up the
line.
3. The needle moves upwards out of the needle cleaning container.
4. Remove the needle from the holder and insert it into the prime bottle.
5. Open the load and unload drawers completely and remove all racks.
HistoCore SPECTRA CV
(→ P. 32 – 4.6 Refilling consumables)
Module Status
and press the
Cleaning and Maintenance
3
2
1
Fig. 88
(→ P. 18 – 3.1 Standard delivery –
Prime/Clean
button of the respective
(→ P. 81 –
must be observed when
L1
or
L2
7
coverslip
111