6.3.6 Load drawer

Warning
The warning messages in
solvents like xylene!
Fill level check of the reagent vessels in the load drawer
Correct fill level of the reagent vessels
1. To scan the fill level of the reagent vessels, press the drawer button of the load drawer
(→ Fig. 50-1).
2. The load drawer opens.
3. If present, remove the cover for the reagent vessel.
4.
Check that the fill level is sufficient (→ Fig. 67)
contamination, the xylene is to be replaced
and insert it into the load drawer)
Warning
A reagent vessel fill level that is too low can cause losses in quality during coverslipping.
5. A sufficient filling level is ensured if the level of the reagent is between the maximum
(→ Fig. 67-1)
HistoCore SPECTRA CV
1
2
Fig. 66
(→ P. 32 – 4.6 Refilling consumables)
and the reagent vessel is to be cleaned before filling it again.
and the minimum (→ Fig. 67-2) fill level mark.
Daily Instrument Setup
2
3
must be observed when handling
and the reagent is free of dirt. In the event of
(→ P. 37 – 4.6.3 Prepare the reagent vessel, fill it
6
87