9. Close the hood.
10. The mounting medium bottles that are reinserted are detected by the instrument software and
the
Module Status
Note
The CMS detects that a new mounting medium bottle has been inserted, and informs the user that
an Extended Prime is required for the coverslip line
means, air and any old mounting medium that is possibly still present is ousted from the hose system.
Only afterwards is the instrument ready to start.
6.3.2 Monitoring and refilling of the needle cleaning container
Note
The HistoCore SPECTRA CV has an automatic fill level scan mechanism in the needle cleaning
container. A warning message informs the user if there is not a sufficient quantity of xylene left in
the glass vessel. Despite the automated fill level scan, Leica recommends also carrying out a visual
inspection during daily startup to ensure that no refilling is necessary during daily routine work.
Warning
The warning messages in
solvents like xylene!
1. Select the
the
Prime/Clean
2. The needle moves automatically to the maintenance position.
3. Open the hood.
4. Take the needle out of the holder sideways and place it in the prime bottle (→ Fig. 3-11).
5. Turn the knurled screw
6. The needle cleaning container
(→ Fig. 59-3).
7. Dispose remaining xylene in the needle cleaning container according to the laboratory
regulations.
8. Outside of the instrument, use a Pasteur pipette to fill xylene into the glass cylinder
(→ Fig. 59-4)
9. Following this, put the needle cleaning container back into the correct position within the
instrument once again and push it completely in.
10. Take the needle out of the prime bottle and reinsert it into the holder.
11. In the Maintenance menu, press the
12. Close the hood.
13. The needle is automatically immersed in the needle cleaning container.
HistoCore SPECTRA CV
(→ Fig. 53)
is updated.
(→ P. 32 – 4.6 Refilling consumables)
Module Status
(→ P. 74 – 5.11 Module status)
button in the desired coverslip line.
(→ Fig. 59-1)
(→ Fig. 59-2)
up to the edge of the plastic cap (approx. 10 ml).
L2
(→ P. 122 – 7.3.2 Extended
must be observed when handling
menu in the main menu and press
clockwise to lift the needle cleaning container.
is lifted upwards and can be removed
Close
button.
Daily Instrument Setup
Prime). By this
6
81