Performance Verification

Spectrum Analyzer Absolute Amplitude Accuracy

The tests in this section verify the absolute amplitude accuracy of the MS2723B or MS2724B Spectrum
Analyzer. This test has three parts:
"50 MHz Amplitude Accuracy Verification."
"Amplitude Accuracy Across Frequency Verification"
"9 kHz to 100 kHz Amplitude Accuracy Verification."
Required Equipment:
• Anritsu MG3692x Synthesized Signal Generator
• Frequency Reference Symmetricom Rubisource T&M
• Power Meter, Anritsu Model ML2438A
• (2) Power Sensor, Anritsu Model MA2442D
• Power Sensor, Anritsu Model MA2421D or Power Sensor, Model SC7816
• Power Splitter, Aeroflex⁄Weinschel Model 1870A
• Fixed Attenuator, Aeroflex/Weinschel Model 44-10
• Spectrum Master: MS2723B or MS2724B
• (2) Adapter, Anritsu Model 34NN50A
• Adapter, Dual Banana Plug to BNC(f)
• Adapter, BNC to BNC
• Coaxial Cable, BNC(m) to BNC(m), 50 ohm
• Low Frequency Cal Fixture, Anritsu Model T3449