2-2 Spectrum Analyzer Function Verification
78. Press the Freq key and press the Start Freq soft key.
79. Use the keypad to enter 14719, and press the MHz soft key.
80. Press Stop Freq soft key, enter 14721, and press the MHz soft key.
81. Press the Shift key and then press the Trace (5) key. Press the Trace A Operations soft key.
82. Press the # of Averages soft key, use the keypad to enter 5, and then press the Enter key.
83. Wait until the Trace Count displays 5/5.
84. Press the Marker key and press the Peak Search soft key.
85. Record the Marker 1 amplitude reading into the "Measured Spur Level" column of
86. Calculate the input related spurious level by subtracting the Marker 1 reading in
Marker 1 reading in
87. Verify that the calculated result is
2-20
Step
85. Record the result into the "Calculated IRS" column of
≤
–60 dBc.
PN: 10580-00165 Rev. D
Performance Verification
Table
A-17.
Step 77
from the
Table
A-17.
Models MS2723B and MS2724B MM
Need help?
Do you have a question about the Spectrum Master MS2723B and is the answer not in the manual?