2-2 Spectrum Analyzer Function Verification
5. Connect the power sensor to the splitter output and adjust the power of the MG3692x so that –20 dBm
appears at the splitter output (approximately +16 dBm).
6. Turn Off the RF of the MG3692x Synthesized Signal Generator.
7. On the Agilent 8648D Signal Generator, turn the RF On, and adjust the RF amplitude so that –20 dBm
appears at the splitter output (approximately +16 dBm).
8. Set the MS2723B or MS2724B to Spectrum Analyzer mode and then preset the instrument.
9. Set the MS2723B or MS2724B as follows:
Center Frequency: – 2400.151 MHz
Span: 100 Hz.
RBW: 30 Hz
VBW: 1 Hz
Reference Level: –15 dBm
Atten Lvl: 0 dB
Preamp: Off
10. Press the Shift key and the Sweep (3) key. Press Detection, and press RMS (red dot appears on the label).
11. Disconnect the power sensor from the output of the splitter and connect the splitter output to the
MS2723B or MS2724B Spectrum Analyzer RF In connector through the Anritsu Model 34NN50A
adapter.