Incorrect reference wavelength selected
Select a different reference wavelength. For guidelines to follow when
selecting the reference wavelength, see
Bandwidth too wide
Select a bandwidth that provides the resolution required to achieve the desired
sensitivity. Narrow spectra peaks require narrow bandwidths. A wider
bandwidth reduces baseline noise, but may reduce peak height.
Reference bandwidth too wide
Select a bandwidth that provides the resolution required to achieve the desired
sensitivity. A wider bandwidth reduces baseline noise, but may increase the
chance that the sample absorbs in the reference bandwidth, thus causing
reduced sensitivity and peak height.
Cell flow is reversed
Check that the flow cell inlet tubing is routed through the heat exchanger (see
Figure
2-5) and is correctly connected to the inside front panel (see
Figure
5-1). Check that solvent flow is directed to the cell inlet and out the
cell outlet.
Semi-micro cell flow rate selected when standard cell is installed
For operation at less than 1.0 mL/min, install a semi-micro flow cell.