PDA is running a calibration or diagnostic function. Not
ready to accept this command.
To troubleshoot:
Wait until the calibration procedure or diagnostic routine finishes running
before attempting to issue any command.
The minimum wavelength must be less than the maximum
wavelength. Min wavelength = x; max wavelength = x.
To troubleshoot:
Select a minimum wavelength that is less than the maximum wavelength.
Both wavelengths must be within the 190 to 800 nm range.
The range of wavelength bunching is 1 to 25 nm.
To troubleshoot:
This message appears if you enter an invalid setting for the bunch width.
Select a bunch width setting from 1 to 25 nm.
The range of wavelengths is outside the 190 to 800 nm limit.
Wavelength = x; Bandwidth = x.
To troubleshoot:
The Dionex PDA can measure the absorbance spectrum from 190 to 800 nm.
Make sure both the minimum and maximum wavelengths are within this
range.