Thermo Scientific Dionex ICS Series Operator's Manual page 72

Photodiode array detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex PDA Operator's Manual
Autozero is already in progress.
To troubleshoot:
This message appears if you select Autozero (in the Chromeleon 7 Command
window or the Chromeleon 6.8 Commands dialog box) when the autozero
routine is already in progress. Allow the autozero routine to continue running
until completion.
Calibration is running.
To troubleshoot:
This message appears if you attempt to issue a command or select a parameter
while a calibration procedure is in progress. Wait until the procedure finishes
running before making any changes.
Cannot change 3DFIELD parameters while data acquisition is
running.
To troubleshoot:
This message appears if you attempt to select a parameter for the 3D field
while data acquisition is in progress. Wait until data acquisition is complete,
or stop the run and then change the parameters.
Cannot set manual parameters for the required operation
because the DSP is not ready.
To troubleshoot:
This message appears if certain commands are issued before the DSP (digital
signal processor) is ready to accept them. Turn off the Dionex PDA power for
30 seconds, and then turn it on again. If the error message appears again,
contact Thermo Fisher Scientific for assistance.
64
Doc. 065378-01 1/12