Thermo Scientific Dionex ICS Series Operator's Manual page 69

Photodiode array detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex PDA-Related Alarms and Error Conditions
The minimum wavelength must be less than the
maximum wavelength. Min wavelength = x; max
wavelength = x.
The range of wavelength bunching is 1 to 25 nm.
The range of wavelengths is outside the 190 to 800 nm
limit. Wavelength = x; Bandwidth = x.
UV lamp control circuitry error while lamp is off.
UV lamp error or control circuitry error while lamp is on.
UV lamp error.
UV lamp error while attempting to turn off.
UV lamp error while attempting to turn on.
UV lamp failed to turn on for the required operation.
Visible lamp control circuitry error while lamp is off.
Visible lamp error or control circuitry error while lamp is
on.
Visible lamp error.
Visible lamp error while attempting to turn off.
Visible lamp error while attempting to turn on.
When collecting at 20 Hz, the spectrum range cannot be
more than 310 nm. Please lower the data rate or narrow
the spectrum range.
Doc. 065378-01 1/12
4 • Troubleshooting
Default Severity
See
Level
Error
page 69
Error
page 69
Error
page 69
Abort
page 72
Abort
page 70
Abort
page 70
Abort
page 70
Abort
page 70
Abort
page 70
Abort
page 72
Abort
page 70
Abort
page 70
Abort
page 72
Abort
page 70
Abort
page 72