B.6 Connecting the Relays/TTLs (Optional)
Connection of relay loads and their power sources to the TTL outputs
will damage the TTL output stage. If the relay load can source more
than 200 mA at 5V or higher, it may damage the Data Processing
board.
SHOCK AND FIRE HAZARD—Inputs and outputs are not rated for
connection to an AC power main. DO NOT CONNECT TO 100, 115, 220,
230, OR 240 VAC POWER SOURCES; FIRE OR ELECTROCUTION MAY
RESULT.
DANGER
d'alimentation électrique fonctionne aux potentiels du secteur. Faites
effectuer toutes les réparations par un personnel qualifié.
STROMSCHLAG
Stromversorgung wird mit Netzspannung betrieben. Der Service darf
nur von qualifiziertem Personal durchgeführt werden.
The Dionex PDA Ship Kit (P/N 060977) includes a 12-pin connector
(P/N 923686) for relay and TTL connections (see
individual wires and twisted pairs to the connectors, strip the end of the wire(s),
insert into the connector, and tighten the locking screw(s) with a small
screwdriver (P/N 046985).
NOTE A twisted wire assembly (P/N 043598) is available as an
Doc. 065378-01 1/12
D'ÉLECTROCUTION
UND
BRANDGEFAHR
option.
RELAY OUT
TTL
OUT
1
2
(+)
1
2
1
2
3
4
5
6
Figure B-4. Relay and TTL Connector
ET
D'INCENDIE
Das
Figure
B-4). To attach
TTL
TTL IN (+)
GND
1
2
3
1
2
4
7
8
9
10
11
12
B • Installation
Le
module
Modul
zur
117

Advertisement