Flow Cells - Thermo Scientific Dionex ICS Series Operator's Manual

Photodiode array detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex PDA Operator's Manual
2.3.1

Flow Cells

Table 2-1
STANDARD FLOW CELLS
Cell Material
PEEK
(P/N 056346)
316 stainless steel
(P/N 056126)
SEMI-MICRO FLOW CELLS
Cell Material
PEEK
(P/N 064169)
316 stainless steel
(P/N 064168)
SEMI-PREPARATIVE FLOW CELL
Cell Material
PEEK
(P/N 064167)
Do not touch the cell windows. If you touch a window, clean it with
isopropyl alcohol (IPA) and a clean lens tissue.
Strong bases can etch the fused silica windows of the flow cell. If the
mobile phase is a base, make sure the mobile phase concentration
does not exceed 0.1 M. If the concentration of the base is greater than
50 mM, disconnect the separator column and flush the system with
deionized water for 5 minutes at 1.0 mL/min immediately after the
analysis. If strong base remains in the cell for 1 to 2 days, the cell
windows may need to be replaced (see
16
lists the flow cells available for use with the Dionex PDA.
Cell
Cell Path
Volume
Length
13 L
10 mm
13 L
10 mm
Cell
Cell Path
Volume
Length
3.1 L
9 mm
3.1 L
9 mm
Cell
Cell Path
Volume
Length
0.7 L
0.4 mm
Table 2-1. Dionex PDA Flow Cells
Volume of Heat
Intended Flow
Exchanger +
Rate Range
Inlet Tubing
45 L
0.5 to 5.0 mL/min
20 L
0.5 to 5.0 mL/min
Volume of Heat
Intended Flow
Exchanger +
Rate Range
Inlet Tubing
5 L
Up to 1.0 mL/min
5 L
Up to 1.0 mL/min
Volume of Heat
Intended Flow
Exchanger +
Rate Range
Inlet Tubing
N/A
5.0 to
100.0 mL/min
Section
5.5).
Doc. 065378-01 1/12