To connect the Dionex PDA to an external USB hub:
The Dionex PDA Ship Kit (P/N 060977) contains one USB cable
(P/N 063246). Contact Thermo Fisher Scientific to order the external
USB hub (P/N 060392) and additional USB cable (P/N 063246) required
for this configuration.
The USB standard limits the USB cable length to 5 meters
(5.5 yds). Each USB device can be separated from the PC by no
more than five hubs. Thus, if five hubs are installed, each USB
device can be located no more than 30 meters (32 yds) from
the PC.
Carefully secure all USB cables, the USB hub, and the hub
power cable so that they cannot be accidentally disconnected.
1. Plug the "A" connector of a USB cable (P/N 063246) into a port on
2. Plug the "B" connector of the cable into the USB receptacle on the
3. Plug the "A" connector of the second USB cable (P/N 063246) into a
4. Plug the "B" connector of the USB cable into a port on the external
PDA
Plug the USB cable's "B" connector
into the USB receptacle on the PDA.
Plug the USB cable's "A" connector
into the "A" port on the USB hub.
122
the external USB hub (P/N 060392) (see
Dionex PDA rear panel.
USB port on the computer on which Chromeleon is installed.
USB hub (see
Figure
USB
Device
B
B
A
USB Hub
Figure B-7. Example USB Connections:
Modules Connected to an External Hub
B-7).
USB
Device
B
USB Cables
A
Plug the USB cable's "A" connector
A
A
into the USB port on the PC.
Plug the USB cable's "B" connector
into the "B" port on the USB hub.
B
Figure
B-7).
PC
Doc. 065378-01 1/12