Photodiode array detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex PDA Operator's Manual
3DFIELD RefWavelength and/or RefBandwidth is out of
range.
To troubleshoot:
For all data collection rates other than 20 Hz, select parameters that meet the
following criteria:
Where:
= Reference wavelength
BW = Reference bandwidth
For example, the following parameters are VALID:
Wavelength = 190 to 800 nm
Reference wavelength = 750 nm
Reference bandwidth = 50 nm
These parameters are valid because 750 nm + 25 nm (50% of 50 nm) =
775 nm, which is less than 800 nm.
The following parameters are NOT VALID:
Wavelength = 190 to 600 nm
Reference wavelength = 750 nm
Reference bandwidth = 50 nm
These parameters are invalid because 750 nm + 25 nm (50% of 50 nm) =
775 nm, which is greater than 600 nm.
62
BW
190
λ
800
2
Doc. 065378-01 1/12