Thermo Scientific Dionex ICS Series Operator's Manual page 61

Photodiode array detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

step determines the sampling rate at which the detector stores spectra.
Ideally, the step is no smaller than the interval suggested for the rise time
(see the table in
larger step size is that it reduces the amount of data stored; for example,
selecting a step twice as large as that listed in the table reduces the data
file size for the 3D field by 50%.
If you select a larger step size when the average parameter is on (see
"Average (Chromeleon 6.8 only)" on page
combination of the selected rise time and the average of the data points
between steps. If you select a larger step size when the average parameter
is off, the selected rise time is valid, but some data points will be skipped.
Average (Chromeleon 6.8 only)
The Average parameter operates in conjunction with the Step parameter (see
(Chromeleon 6.8 only)" on page
When Average is On, Chromeleon 6.8 averages the data points between step
intervals, reports that value, and noise is reduced. This setting is recommended for
most applications.
When Average is Off, Chromeleon 6.8 reports the data points at the step interval.
The data points between steps are skipped and noise is not reduced.
If Average is On and the step size is more than double the rise time,
the system response is dominated by the step size. The system
response time is approximately the larger of (a) the rise time or (b) two
times the step size (when Average is On).
Negative Absorbance
Negative absorbance is the result of decreased baseline absorption of the mobile
phase after the autozero routine (which usually occurs at the beginning of a run).
At the default setting of 1, the Dionex PDA becomes saturated when the light
transmission is 20% above the autozero level. This results in a flat, truncated
baseline.
If desired, increase the negative absorbance level in order to continue seeing the
actual baseline and peaks. This is especially helpful when running a gradient
application in which the absorbance decreases more than 100 mAU. As the
following table indicates, the extent of negative absorbance depends on the
wavelength. (The values in this table are intended as guidelines only.)
Doc. 065378-01 1/12
3 • Operation and Maintenance
"Data Collection Rate" on page
52).
45). The advantage of a
53), the system response is a
"Step
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dionex ad25Dionex pdaDionex pda-100Dionex pda-1074114

Table of Contents