Intervals; Preparing Hygienic Reprocessing - Weinmann LIFE-BASE light XS Instructions For Use Manual

Portable unit
Table of Contents

Advertisement

8 Hygienic reprocessing

8.1 Intervals

8.2 Preparing hygienic reprocessing

Requirement
150
EN
LIFE-BASE
Part
Interval
After
each use*
Portable unit
including
accessories
Protective bags
*In the case of visible soiling or suspected contamination.
The devices attached to the portable unit are switched off.
There is no connection between devices and patient.
1. Disconnect the portable unit from the power supply.
2. If available:
Remove the portable unit from the wall mounting or ceiling
mounting
(see "6.2 Removing the portable unit from the
wall mounting or ceiling mounting", page 92)
or
Remove the portable unit from the pole or equipment rail
(see "6.4 Removing the portable unit from the pole or
equipment rail", page
3. Remove the accessories from the devices.
4. If available: Remove the oxygen cylinder from the portable unit
(see "7.4 Removing the oxygen cylinder from the portable
unit", page
121).
5. If available: Remove the belts from the portable unit
(see "7.3 Removing the belt from the portable unit",
page
119).
At least
1x weekly
103).
At least
1x monthly

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Life-base 1 ngLife-base iiiLife-base lightLife-base 1 ng xlLife-base 3 ngLife-base 1 ng xs ... Show all

Table of Contents