7 Disassembly
Result
7.5.5 Removing the device from LIFE-BASE 1 NG and
Applies to
Required auxiliary
equipment
Requirement
128
EN
LIFE-BASE
6. Remove the spacer plate from the device.
The device is removed from the portable unit.
LIFE-BASE 4 NG (with protective hoops)
MEDUMAT Transport
Screwdriver with crosshead PH1
Disassembly is prepared
page
123).
1. Place the portable unit and device on a flat, firm surface.
2. Undo 4 screws 103, 4 protective disks 104 and 4 plain
washers 106 from the portable unit.
3. Remove the device from the portable unit.
(see "7.5.1 Preparing disassembly",
106
104
103