Fastening The Portable Unit To The Pole Or Equipment Rail - Weinmann LIFE-BASE light XS Instructions For Use Manual

Portable unit
Table of Contents

Advertisement

6.3 Fastening the portable unit to the
pole or equipment rail
Risk of injury due to inadequate fastening of the portable
unit!
If the portable unit is not fastened properly or only suspended
loosely using the hangers or holding claws, the portable unit may
fall off or be thrown around inside the emergency vehicle and
injure people. Attachment by loose suspension, with straps or
pole mounting is not sufficient to hold the portable unit safely
according to EN 1789.
 Fasten the portable unit to a wall mounting, ceiling mounting
or an equipment rail using retaining claws during travel or
flight.
 Only fasten the portable unit with retaining claws to
equipment rails which are designed to hold a weight of at least
6 kg.
6.3.1 Fastening LIFE-BASE Mini II
Applies to
LIFE-BASE Mini II
1. Suspend the back of the portable unit on the pole or
equipment rail.
Result
LIFE-BASE Mini II is fastened on a pole or an equipment rail.
6 Operation
LIFE-BASE
EN
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Life-base 1 ngLife-base iiiLife-base lightLife-base 1 ng xlLife-base 3 ngLife-base 1 ng xs ... Show all

Table of Contents