Main Screen; O 2 Monitoring; Parameter Field For O 2 Monitoring - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Main screen

 Press the
(A) key.
O
monitoring
2
The inspiratory oxygen concentration is measured
with a dual galvanic sensor, which is located in the
cover of the inspiratory valve, see chapter
''Compact breathing system COSY (top view)'' on
page 21.
CAUTION
Risk of inaccurate measured values
If the O
sensor is removed, this can lead to
2
leaks in the breathing system.
When the O
sensor is replaced or removed, it
2
must be recalibrated.
NOTE
If the anesthesia workstation is not used, remove
the O
sensor from the cover of the inspiratory
2
valve and insert the sealing plug provided.
Instructions for use Fabius Tiro SW 3.n
The current dialog window changes to the main
screen.
The following information is displayed on the
screen:
A
– Current alarm messages
– Data of the O
– Data of the airway pressure monitoring
– Data of the breathing volume monitoring
Parameter field for O
Volume Control
The following information is displayed in the
parameter field for O
A Numerical value for the inspiratory O
concentration in percent (%) in the range from
10 % to 100 %
B Upper alarm limit for the O
in (%)
C Lower alarm limit for the O
in (%)
Monitoring
monitoring
2
monitoring
2
A
monitoring:
2
concentration
2
concentration
2
B
C
2
117