Piping Parts For A System Using Detector With Pressure Compensation4-14; Piping For The Calibration Gas; Piping For The Reference Gas - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

4.4.1
Piping Parts for a System using Detector with Pressure Compensation
Detector
Detector with
pressure
compensation
Note:
Parts with marking * are used when required.
General parts can be found on the open market.
4.4.2

Piping for the Calibration Gas

4.4.3

Piping for the Reference Gas

4-14
Check that the parts listed in Table 4.3 are ready.
Table 4.3
Piping location
Calibration gas inlet
Stop valve or check valve
Nipple *
Zero gas cylinder
Gas pressure regulator
Joint for tube connection
Reference gas inlet
Air set
Joint for tube connection
Calibration gas piping is basically identical to that of System 2. See Section 4.2.2.
Reference gas piping is basically identical to that of for System 2. See Section 4.2.3.
Parts
Recommended by YOKOGAWA (L9852CB or
G7016XH)Provided by YOKOGAWA (K9292DN
or K9292DS)
Rc1/4 or 1/4 NPT
User«s scope
of a quality specified by YOKOGAWA (G7013XF
or G7014XF)
Rc1/4 or 1/4 NPT
of a quality specified by YOKOGAWA (G7011XF
or E7040EL)
Rc1/4 or 1/4 NPT
Note
general parts
general parts
general parts
T4.3E.EPS
IM 11M12A01-02E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22gZr402g

Table of Contents