Preference Order Of Output Hold Value - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

8.2.2

Preference Order of Output Hold Value

IM 11M12A01-02E
For semi-automatic calibration, "under calibration" is the time required from entering
calibration instructions to perform a, either by using the touchpanel or by a contact
input, calibration until the output stabilization time elapses.
For automatic calibration, "under calibration" is the time required, after performing an
appropriate calibration until the output stabilization time elapses.
(4) During "Blow-back" (see Section 10.2, "Blow-back," later in this manual)
During semi-automatic blow-back:
"During blow-back" is the time required after pressing the blow-back start key, by using
the touchpanel or entering a blow-back start instruction by using a contact input, until
the blow-back time and output stabilization time elapse.
During automatic blow-back:
"During blow-back" is the time required after reaching the blow-back start time until the
blow-back time and output stabilization time elapse.
(5) On Error occurrence
This is the time at which any of Errors 1 to 4 occurs.
The output hold value takes the following preference order:
Preference order (high)
For example, if the output current is set to 4 mA during maintenance, and no output-
hold output for during calibration is preset, the output is held at 4 mA during the
maintenance display. However, the output hold is released at the time of starting the
calibration, and the output will be again held at 4 mA after completing the calibration
and when the output stabilization time elapses.
On error occurrence
Under calibration or during blow-back
During warm-up
Under maintenance
8. Detailed Data Setting
8.2.2.siki
8-5