7.11 Checking Contact I/O

7.11.1 Checking Contact Output

7-12
Conduct the contact input and output checking as well as operational checking of the
solenoid valves for automatic calibration.
To check the contact output, follow these steps:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Select Maintenance in that display.
(2) Select Contact check then contact output in the Maintenance panel display to display
the Output contact check display (see Figure 7.15.2).
(3) In this display, select the desired output contact for checking. The display, which
enables the closing and opening of contacts, then appears. Use the display to conduct
continuity checking.
Output contacts
Output contact 1 : Open
Output contact 2 : Open
Output contact 3 : Open
Output contact 4 : Open
Figure 7.15.2 Output Contact Check Panel Display
CAUTION
• If you conduct an open-close check for contact output 4, Error 1 or Error 2 will occur.
This is because the built-in heater power of the detector, which is connected to
contact output 4, is turned off during the above check. So, if the above error occurs,
reset the equipment or turn the power off and then back on to restart (refer to Section
10.4, "Resetting," later in this manual).
Enter
F7.15.2E.EPS