Setting Procedure; Default Values - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

8.5.2

Setting Procedure

8.5.3

Default Values

IM 11M12A01-02E
The following are set so that semi-automatic calibration starts when input contact open
is applied to "Input1."
Proceed as follows:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.
(2) Select Setup in the Execution/Setup display to display the "Commissioning" (Setup)
display.
(3) Select "Contact setup" in the "Commissioning" (Setup) display.
(4) Select "Input contacts" in the Contact setup display. The Input contacts display then
appears, as shown in Figure 8.11.
(5) Select Input1 in that display. The Input contacts display then appears, as shown in
Figure 8.12.
(6) Select Calibration start.
(7) Select Input1 "closed." An Open or Closed selection display then appears.
(8) Choose Open.
Input contacts
Input1: No function
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Figure 8.11 Input Contacts Display
When the analyzer is delivered, or if data are initialized, contact input is diaable.
Input contacts
Input1: No function
Input1 contact is active at:
Input2: No function
Input2 contact is active at:
Enter
F8.11E.EPS
Figure 8.12 Input Contacts Display
8. Detailed Data Setting
No function
Cal. gas press. low
Range change
Closed
Calibration start
Process up set
Blow back start
Closed
Enter
F8.12.EPS
8-15