8.2 Output Hold Setting

8.2.1

Definition of Equipment Status

8-4
The "output hold" functions hold an analog output signal at a preset value during the
equipment's warm-up time or calibration or if an error arises. Outputs 1 and 2 can be set
individually. Table 8.1.2 shows the analog outputs that can be retained and the indi-
vidual states.
Table 8.1.2
Equipment status
During warm-up
Output hold
values available
4 mA
20 mA
Without hold
feature
Retains output
from just before
occurrence
Set value
(2.4 to 21.6 mA)
: The output hold functions are available.
(1) During warm up
"During warm up" is the time required after applying power until the sensor temperature
stabilizes at 750 C, and the equipment is in the measurement mode.
(2) Under maintenance
"Under maintenance" is the time, starting for pressing the Setup key in the basic panel
display the Execution/Setup display, to go back to the basic panel display. The display
panels under the Execution/Setup display, which are in Figure 7.3.1, are included.
(3) "Under calibration" (see Chapter 3, Calibration)
"Under calibration" is the time, during manual calibration, starting from entering the
Manual calibration display (shown in Figure 8.1), and making a series of calibrations,
until the Calibration End key is pressed or until the preset output stabilization time
elapses.
Manual calibration
Open span gas valve.
Set flow span gas to
600ml/min.
Valve opened
Cancel calibration
Figure 8.1
Under
Under calibration
maintenance
During blow-back
Enter
F8.1E.EPS
On Error
occurrence
T8.1.2E.EPS
IM 11M12A01-02E