10.3 Operational Data Initialization

IM 11M12A01-02E
Individual set data initialization enables you to return to the default values set at the time
of delivery. There are two types of initializations: an all set-data initialization and a
function-by-function initialization. Table 10.5 lists the initialization items and default
values.
To initialize the set data, follow these steps:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.
Then choose Setup.
(2) Select Others from the "Commissioning" (Setup) display.
(3) Choose "Defaults." The "Defaults" display shown in Figure 10.11 appears.
(4) Select your desired item to initialize in order to display the "Defaults Display"
shown in Figure 10.12.
(5) Choose Defaults start. Initialization then starts.
Defaults
Display
Calibration
Blow back
mA-outputs
Alarms
Contacts
Others
All data
Figure 10.11 Defaults Display
WARNING
• Do NOT turn off the power during initialization. Otherwise, initialization will not be
performed properly.
Defaults
Defaults start
Cancel
Enter
F10.11E.EPS
Figure 10.12 Initialization Start Display
10. Other Functions
Display
Enter
10-11