Setting Display Item - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

7.9 Setting Display Item

IM 11M12A01-02E
This section briefly describes the display item settings shown in Figure 7.11, "Basic
Panel Display."
Tag name
Primary value
Secondary value
Tertiary value
Figure 7.11 Basic Panel Display
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Then select Maintenance in the Execution/Setup display.
(2) Select the Display setup from the Maintenance panel display (Figure 7.12). The
Display setup display (Figure 7.13) then appears.
(3) In the above Display setup display, select the Display item. The Display item display
(Figure 7.14) then appears. From this display, select the Primary value and press the
[Enter] key to display the Display item selection display (Figure 7.15).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, "Display Items," enabling the selection of display items in
individual display areas.
Maintenance
Display setup
Calibration setup
Blow back setup
mA-output loop check
Contact check
Figure 7.12 Maintenance Panel Display
Display item
Primary value:
Oxygen
Secondary value:
mA-output1
Tertiary value:
mA-output2
Tag name:
Figure 7.14 Display Item Display
Tag:
21.0
17.43mA -Output1
17.43mA -Output2
Display item
Trend graph
Auto return time:
Language:
Enter
F7.12.EPS
Figure 7.13 Display Setup
Primary value:
Oxygen
Oxygen
Item of output1 damping
Secondary value:
Item of output2 damping
mA-output1
Tertiary value:
mA-output2
Tag name:
Enter
F7.14.EPS
Figure 7.15 Display Item Selection
%O
2
F7.11E.EPS
Display setup
0min
English
Display item
7. Startup
Enter
F7.13.EPS
Enter
F7.15.EPS
7-9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22gZr402g

Table of Contents