Zr402G Converter - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

6.2 ZR402G Converter

Complete Operation Display
Interactive operations along with operation
display
A variety of display modes enabling you to
select the operation mode freely
Back-lit LCD display allows viewing even in areas
of low lighting
Error codes and details of errors are displayed no
need to refer to instruction manual
Password for security
Self-testing suggests countermeasures for
problems
If a problem occurs, the liquid-crystal display will
provides an error code and description of the problem.
This enables prompt and appropriate corrective action
to be taken.
Error code
E--1
Cell failure
E--2
Abnormal heater temperature
E--3
Defective A/D converter
E--4
Faulty EEPROM
ALARM1
Abnormal oxygen concentration
ALARM2
Abnormal moisture
ALARM3
Abnormal mixing ratio
ALARM6
Abnormal zero calibration factor
ALARM7
Abnormal span calibration factor
ALARM8
Stabilization time over
IM 11M12A01-02E
Reason for error
Figure 6.3 Converter
Typical Converter Displays
Example of basic display
This data display provides for interactive operation.
Example of trend display displays data
changes
During automatic calibration,you can check has stabilized,
display data while viewing oxygen trend data,thus providing
highly reliable calibration.
Example of setting data display displays data
changes
Commissioning
Basic setup
mA-output setup
Alarms setup
Contact setup
Others
One-touch interactive display operation
User-friendly design providing easy operation
without need to use the instruction manual