Installation Of The Detector - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

100 mm
(Note)
When using the detector with pressure compensation, ensure that the flange
gasket does not block the reference air outlet on the detector flange. If the flange
gasket blocks the outlet, the detector cannot perform pressure compensation.
Where necessary, make a notch in the flange gasket. Confirm the outside
dimensions of the detector in Chapter 3.8 before installation.
3.1.3

Installation of the Detector

CAUTION
3-2
(1) Do not mount the probe with the tip higher than the probe base.
(2) If the probe length is 2.5 meters or more, the detector should be mounted vertically
(no more than a 5
tilt).
(3) The detector probe should be mounted at right angles to the measurement gas flow or
the probe tip should point downstream.
Figure 3.1 illustrates an example of the probe insertion hole.
Bounds of the probe
(vertical)
insertion hole location
*1
Note
Figure 3.1 Example of forming probe insertion hole
• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop
the detector, as impact will damage it.
• A gasket should be used between the flanges to prevent gas leakage. The gasket
material should be heatproof and corrosion-proof, suited to the characteristics of the
measured gas.
The following should be taken into consideration when mounting the general-use
detector:
<General-use detector>
(1) Make sure that the cell mounting screws (four) at the probe tip are not loose.
If a dust filter (see Section 2.8.1) is used, make sure it is properly attached to the
detector. Refer to Section 3.1.4 for installation of the dust filter.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the refer-
ence gas inlet should face downward.
Flange matches
the detector size
*1
Type
Standard
With dust filter
With probe protector
(horizontal)
100 mm
Four-hole flange
Outside diameter of detector
50.8 mm in diameter (Note)
51 mm in diameter (Note)
60.5 mm in diameter (Note)
Eight-hole flange
F3.1E.EPS