Output Hold Setting; Default Values - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

8.2.3

Output Hold Setting

8.2.4

Default Values

8-6
To set the output hold, follow these steps:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Then select Setup in the Execution/Setup display. Next, select the mA-output setup
and then the mA-output preset display as shown in Figure 8.2.
mA-outputs presets
Warm up:
4mA
Preset value:
4 . 0 mA
Maintenance: Hold
Preset value:
4 . 0 mA
Cal.blowback: Hold
Preset value:
4 . 0 mA
Error:
Hold
Preset value:
4 . 0 mA
Figure 8.2 mA-output Preset Display
(2) From this display (Fig. 8.2), select the desired display. Figure 8.3 shows an example
of selecting Maintenance. Select the desired output status.
(3) If a preset value is selected, set the corresponding output current. If you select a
preset value just below Maintenance on the screen, the numeric-data entry display
will appear. Enter the current value you want. To set 10 mA, type in 010 and press
the [Enter] key to complete the setting. The setting range is from 2.4 to 21.6 mA.
When the analyzer is delivered, or if data are initialized, output hold is the default as
shown in Table 8.2.
Table 8.2 Output Hold Default Values
Status
Under warm-up
Under maintenance
During calibration or during
blow-back
On Error occurrence
Set value: all 4 mA
mA-outputs presets
Warm up:
Preset value:
Maintenance: Hold
Preset value:
Cal.blowback: Hold
Preset value:
Error:
Enter
Preset value:
F8.2E.EPS
Figure 8.3
Output hold (min. and max. values)
4 mA
Holds output at value just before maintenance started.
Holds output at value just before starting calibration
or "blow-back."
Holds output at value just before error occurrence.
4mA
4 . 0 mA
Non-Hold
Hold
4.0mA
Preset
4 . 0 mA
Hold
4 . 0 mA
Enter
F8.3E.EPS
T8.2E.EPS