CAUTION
5.1.1

Terminals for the External Wiring in the Converter

Front door
5-2
• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3)
grounding (the grounding resistance is 100
• Special consideration of cable length should be taken for the HART communication,
For the detail, refer to Section 1.1.2 of the IM 11M12A01-51E " Communication Line
Requirement ".
Open the front door and remove the terminal cover to gain access to the converter
external wiring terminals (see Figure 5.2).
Figure 5.2 Terminals for external wiring in the converter
or less).
Terminals are visible
when the terminal cover is removed.
Cable inlet
F5.2E.EPS
IM 11M12A01-02E