Characteristic Data From A Sensor Measured During Calibration - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

9.1.4

Characteristic Data from a Sensor Measured During Calibration

9-4
81.92
ez
Cell
electromotive
Calibration curve before
force, mV
correction
e1
es
C
0
21.0
p1
Span-gas concentration
Oxygen concentration (percent oxygen by volume)
Zero-point correction factor = (B/A) x 100 (percent) Correctable range: 100 ± 30 percent
Span correction factor = (C/A) x 100 (percent)
Figure 9.3 Calculation of a One-point Calibration Curve and Correction Factors
using a Span Gas
During calibration, calibration data and sensor status data (listed below) are acquired.
However, if the calibration is not properly conducted (an error occurs in automatic or
semi-automatic calibration), these data are not collected in the current calibration.
These data can be observed by selecting the detailed data display key from the basic
panel display. For an explanation and the operating procedures of individual data,
consult Section 10.1.1, "Detailed Display."
(1) Record of span correction factor
Recorded the past ten span correction factors.
(2) Record of zero correction factors
Recorded the past ten zero correction factors.
(3) Response time
You can monitor the response time provided that a two-point calibration has been done
in semi-automatic or automatic calibration.
(4) Cell's internal resistance
The cell's internal resistance gradually increases as the cell (sensor) deteriorates. You
can monitor the values measured during the latest calibration. However, these values
include the cell's internal resistance and other wiring connection resistance. So, the
cell's degrading cannot be estimated from these values only.
When only a span calibration has been made, these values will not be measured, and
previously measured values will remain.
(5) Robustness of a cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is
expressed in a number on four levels.
Previous
zero-gas data
Corrected calibration curve
(theoretical calibration curve)
0.51
Correctable range: 0 ± 18 percent
Zero origin
B
A
Span origin
F9.3E.EPS
IM 11M12A01-02E