Default Values - YOKOGAWA EXAxtZR Instruction Manual

Separate type zirconia oxygen analyzer
Table of Contents

Advertisement

8.3.4

Default Values

IM 11M12A01-02E
Alarms setup
Parameter:
Oxygen
Hysteresis:
0.1%O2
Contact delay:
3 s
Setpoints
Figure 8.5 Alarms Setup Display
When the analyzer is delivered, or if data are initialized, the default alarm set values are
as shown in Table 8.3.
Table 8.3 Alarm Setting Default Values
Set item
Hysteresis
Delay time
High-high limit alarm
High-high-limit alarm setpoint
High-limit alarm
High- and low-limit alarm setpoints
Low-limit alarm
Low-limit alarm setpoint
Low-low-limit alarm
Low-low-limit alarm setpoint
Oxygen alarms
High High alarm: OFF
Set value: 1 0 0 . 0 % O2
High alarm:
Set value: 1 0 0 . 0 % O2
Low alarm:
Set value:
Low Low alarm:
Set value:
Enter
F8.5E.EPS
Figure 8.6 Oxygen Alarms Display
Set value
0.1% O
2
30 seconds
Off
100% O
2
Off
100% O
2
Off
0% O
2
Off
0% O
2
T8.3E.EPS
8. Detailed Data Setting
ON
OFF
0 . 0 % O2
OFF
0 . 0 % O2
Enter
F8.6E.EPS
8-9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22gZr402g

Table of Contents