7.10 Checking Current Loop

IM 11M12A01-02E
The set current can be output as an analog output.
(1) Press the Setup key on the basic panel display to display the Execution/Setup
display. Then select Maintenance in the Execution/Setup display.
(2) Select "mA-output loop check" in the Maintenance panel display to display the "mA-
output loop check" display, enabling you to check "mA-output1" and "mA-output2."
Select the desired output terminal for current-loop checking (see Figure 7.15.1).
(3) At the time of entering the numeric-data entry display, the output current will change
to 4 mA (default value). If the desired current is entered, the corresponding output
will be provided.
mA - output loop check
mA - output 1
mA - output 2
Figure 7.15.1 "mA-output loop check" Panel Display
Enter
F7.15.1E.EPS
7. Startup
7-11