6.8 Manual calibration procedure
6.8.1 Manual zero calibration
It is used for zero point adjustment. Proper
zero gas, suitable for the application, should
be used. Refer to Section 3.4.3 "Preparation of
standard gas".
ZERO
(1) Press the
screen to display the Manual Zero Cali-
bration screen.
(2) Select the Ch (component) to be calibrat-
ed by pressing the
selection, press the
gas will be supplied.
CAUTION
For the Ch (components) in which "at once" is
set in the zero calibration (see Section 6.2.2) -
zero calibration is carried out simultaneously.
And for the Ch (components) in which "both"
is set in the calibration range setting
(see Section 6.2.3) - zero calibration is carried
out on both ranges.
Attention
Pour les canaux Ch (composants) dont le
choix de paramétrage est «at once»
(voir Section 6.2.2), le calibrage du zéro se
fera simultanément.
Pour les canaux Ch (composants) dont le
choix de paramétrage est «both»
(voir Section 6.2.3), le calibrage du zéro se
fera simultanément sur les deux échelles.
key on the Measurement
or the
key. After
ENT
key, and zero
62
Measurement Mode Screen
ZERO
(
ENT
)
INZ-TN2ZPA-E