Forced Run/Stop Of Auto Calibration - Fuji Electric ZPA Instruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPA:
Table of Contents

Advertisement

6.4.2 Forced run/stop of auto calibration

Auto calibration can be performed just once or forcibly stopped while the calibration is performed.
6.4.2.1 Execution of auto calibration (only once)
(1) In the "Setting of Auto Calibration"
screen that appears, point the cursor to
"Auto Calibration Run" by pressing the
or the
(2) "Run" is highlighted, displaying a mes-
sage to confi rm the execution of auto
calibration. Press the
the auto calibration, and press the
key to cancel.
6.4.2.2 Forced stop of auto calibration
This mode is used to stop the auto calibration forcibly.
(1) In the "Setting of Auto Calibration"
screen that appears, point the cursor to
"Auto Calibration Stop" by pressing the
or the
("Auto Calibration Stop" appears when
the screen is selected while auto calibra-
tion is performed.)
(2) "Stop" is highlighted, displaying a mes-
sage to confi rm the stop of auto calibra-
tion. Press the
calibration, and press the
cancel (not stopped).
ENT
key. Press the
key.
ENT
key to execute
ESC
ENT
key. Press the
key.
ENT
key to stop the auto
ESC
key to
44
INZ-TN2ZPA-E