6.5.2 Forced run/stop of auto zero calibration

Auto zero calibration can be performed just once, or auto zero calibration can be forcibly stopped
during calibration.
6.5.2.1 Execution of auto zero calibration (only once)
(1) In the "Setting of Auto Zero Calibration"
screen that appears, point the cursor to
"Run" by pressing the
ENT
Press the
(2) "Run" is highlighted, displaying a mes-
sage to confi rm execution of auto zero
calibration. Press the
the calibration, and press the
cancel.
6.5.2.2 Forced stop of auto zero calibration
This mode is used to cancel the auto zero calibration forcedly.
(1) In the "Setting of Auto Zero Calibration"
screen that appears, point the cursor to
"Stop" by pressing the
Press the
ENT
("Auto Zero Calibration Stop" appears
when the screen is selected while auto
zero calibration is performed.)
(2) "Stop" is highlighted, displaying a mes-
sage to confi rm the stop of auto zero cali-
bration. Press the
auto zero calibration and the
cancel (not stopped).
INZ-TN2ZPA-E
or the
key.
key.
ENT
key to execute
ESC
key to
or the
key.
key.
ENT
key to stop the
ESC
key to
49