3. INSTALLATION

DANGER
This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fi re or other serious accidents.
CAUTION
• Entrust the installation, movement or re-installation to a specialist or the supplier. A poor
installation may cause accidental tipover, electric shock, fi re, injury, etc.
• The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tipover
or drop, for example, causing accident or injury.
• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an
injury.
• This unit should be installed in a place which conforms to the conditions noted in the in-
struction manual. Otherwise, it may cause electric shocks, fi re or malfunction of the unit.
• During installation work, care should be taken to keep the unit free from entry of cable chips
or other foreign objects. Otherwise, it may cause fi re, trouble or malfunction of the unit.

3.1 Installation conditions

To install the analyzer for optimum performance, select a location that meets the following
conditions;
(1) This instrument is system built in type. This instrument should be used while embedded in a
panel, locker, or enclosure of steel sheet.
Keep a minimum clearance of 10 cm above the analyzer for heat dissipation. The same
clearance is required for each analyzers when you install several units on a multistage rack.
(2) Use this instrument indoors.
(3) A vibration-free place
(4) A place which is clean around the analyzer.
(5) Power supply
Rated voltage
Operating voltage
Rated frequency
Power consumption
INZ-TN2ZPA-E
: 100V to 240V AC
: 85V to 264V AC
: 50/60 Hz
: 100 VA max.
5