6.1.2 Manual range switch

The range of the measured component can be
switched manually as follows.
(1) Select "MR" as range switch mode, and
then press the
(2) Move the highlight of the cursor to range
selection, and then select a desired range
by pressing the
mark indicates the currently selected
range.)
(3) Then press the
surement is carried out in the selected
range.
Note) If "RR" or "AR" is selected as range
switch mode, this operation cannot
be performed.
The ranges for O
O
correction average value, and
2
O
average value are automatically
2
switched according to the instanta-
neous value range switch settings.
(Same as for "RR" or "AR".)
To close the setting
ESC
Press the
switch mode or range switch operation or stop
the operation in the middle. The setting
operation is made invalid and the previous
screen appears.
Range identification contact operation
The range identification contact output corresponding to each Ch (component) is closed when
Range 1 is active, and open when Range 2 is active, no matter.
If the measurement value is held by remote contact input or during calibration routine and
range switch conditions are met, the contact will change position only after the hold codition
is removed.