CAUTION
Damage to the device
For the combustion of solid fuels (wood / pellets / wood chips), a fil-
tering device recommended by ecom must always be used (see il-
lustration)!
Filter plate
3. Carry out gas analysis
If the calibration has been successfully completed and all previous queries
have been answered, the instrument is ready for operation.
NOTE
The following minimum times must be observed in order to obtain
correct measured values:
1 minute for fresh air calibration of the sensors
2 minutes for stable measured values at the instrument
To obtain accurate measurement values it's important to re-calibrate the in-
strument after each measurement (at the latest after one hour). To calibrate
the sensors, proceed as follows:
• switch off the instrument and wait for the switch off phase
• switch on the device and refer to chapter "Measurement Preparations"
Page 20
Washing bottle
Probe prefilter
ecom-EN3
Need help?